Mapping fundamentals, electrograms and entrainment
Raja Selvaraj, JIPMER
Basics of recording electrograms in the lab
Schematic of EP lab setup
Multi-electrode Catheters
2 - 20 electrodes
Quadripolar
Decapolar
Decapolar catheter - which is electrode 1 ?
Interelectrode spacing
Varying inter-electrode distances
Accurate timing information with 2 or 5 mm distance
Usually 5 mm between electrode pairs
Connector
Junction box
Provides an interface between the catheters and the recording system
Numbered poles - can be used for setting up channels for recording / stimulation
Junction box
Setting up the amplifier
Amplifier
Signals are extremely small
Amplification required
Digitization
Filtering
Digitization
Range (intracardiac signals not more than 10 mV)
Bit resolution
Sampling rate
Digitization
Filters
High pass
Low pass
Notch
High pass filter
Higher high pass limits view to 'local' events
Increasing further reduces the energy of recording
Removes baseline wander and other low frequency noise
Low pass filter
No significant components beyond about 300 Hz
Reduces high frequency noise component
Setting up filters
Frequency content of the signal
High - content up to 300 Hz
Low - T waves
Noise to be avoided
Electrical interference (50 / 60 Hz)
Myopotential / high freq artifacts
Respiration / Baseline wander
ECG
All 12 leads
0.1/0.5 - 50/100 Hz
Bipolar
Adjacent electrodes
30/50 Hz - 300/500 Hz
Notch filter
Unipolar
Intracardiac / WCT for reference electrode
Exploring electrode to positive terminal
DC/0.05 Hz - 300/500 Hz
Amplifier setting
Setting up the display
Pages
12 lead
IC
Abl, others
Intracardiac
3-4 ECG leads, usually orthogonal
Intracardiac
CS display conventions
Colors
Ablation
Essential signals only
Unipolar EGM from mapping catheter
No clipping of ablation signals
Equal gain of proximal and distal ablation signals
Electrograms
Bipolar
Potential difference between two electrodes
Both in contact with myocardium
Usually closely spaced
Bipolar
Rapid, high frequency
Reflects "local" events
Unipolar
Potential difference between "exploring" and "indifferent" electrodes
Exploring electrode is in contact with heart
Indifferent electrode is at distance
WCT
Electrode in IVC
Unipolar
Records both local and distant events
Inverse square law
Frequency higher for local events
Local activation time
Bipolar electrogram
Intrinsic deflection
Zero crossing
No indication of timing in relation to origin
Unipolar electrogram
Maximum negative dV/dt
Electrograms - accessory pathway
Electrograms - VT
Entrainment
Basic principles
Prerequisites for entrainment
Reentrant circuit
Excitable gap
No entrance block
Identifying entrainment
Constant fusion
Progressive fusion
Only two criteria originally described by Waldo
Sometimes probabilistic
constant PPI at different pacing rates
Interpretation
Deviation from morphology indicates extent of capture by antidromic wavefront
Deviation from cycle length indicates distance from circuit
PPI
Scenario 1 - Atrial flutter
Pacing CL
Pace at 20-30 ms shorter than TCL
Faster - decrement
Slower - difficult to measure
How to do
Stim set up
Atrial activation sequence
PPI
Interpretation
Identify circuit
Quick locate flutter
Identify isthmus
Catheters
Entrainment
Pacing from lateral RA
Pacing from lateral RA
Pacing from lateral RA
Pacing from lateral RA
Pacing from lateral RA
Pacing from lateral Isthmus
Ventricular tachycardia
Same concept
Constant fusion
Progressive fusion
Example
Example